一、工作原理
1200℃催化(hua)(hua)(hua)燃燒氧化(hua)(hua)(hua)法,將試(shi)樣連同凈化(hua)(hua)(hua)氣(qi)體(高純氧)分(fen)別(bie)導入(ru)高溫(wen)燃燒管(guan)和(he)低(di)溫(wen)反(fan)應(ying)管(guan)中(zhong)(zhong),經高溫(wen)燃燒管(guan)的(de)試(shi)樣被(bei)高溫(wen)催化(hua)(hua)(hua)氧化(hua)(hua)(hua),其中(zhong)(zhong)的(de)有機(ji)(ji)(ji)碳(tan)(tan)和(he)無(wu)機(ji)(ji)(ji)碳(tan)(tan)均轉化(hua)(hua)(hua)為二(er)氧化(hua)(hua)(hua)碳(tan)(tan),經低(di)溫(wen)反(fan)應(ying)管(guan)的(de)試(shi)樣被(bei)酸化(hua)(hua)(hua)后,其中(zhong)(zhong)的(de)無(wu)機(ji)(ji)(ji)碳(tan)(tan)分(fen)解成二(er)氧化(hua)(hua)(hua)碳(tan)(tan),兩(liang)種反(fan)應(ying)管(guan)中(zhong)(zhong)生產(chan)的(de)二(er)氧化(hua)(hua)(hua)碳(tan)(tan)經載氣(qi)輸送依次被(bei)導入(ru)非分(fen)散紅外氣(qi)體檢測(ce)器NDIR中(zhong)(zhong), CO2被(bei)檢測(ce)。從而分(fen)別(bie)測(ce)得水中(zhong)(zhong)的(de)總碳(tan)(tan)(TC)和(he)無(wu)機(ji)(ji)(ji)碳(tan)(tan)(IC)。總碳(tan)(tan)與無(wu)機(ji)(ji)(ji)碳(tan)(tan)之差值,即(ji)為總有機(ji)(ji)(ji)碳(tan)(tan)(TOC)。即(ji):TOC=TC- IC
二、技術參數
1. 測量范圍(0-30000.000)mg/L
2. 測定精(jing)度:CV≤1.5%(重復精(jing)度)
3. TC:0~30000mg/L,檢出限≤50μg/L
4. IC :0~3000mg/L,檢(jian)出限≤4μg/L
5. 分析時間(jian):3min
6. 進樣量 TC:10~150μL, IC:10~450μL,可調節
7. 接口、兼容性及相互配(pei)合:
8. COM1:數據(ju)傳輸
9. COM2:數據控制
10.com3:自動取樣(ASE-18C、ASE-24、ASE-70)
11. com4:5-20Ma